T-Series/Commands/Restore Settings/A-Series

From ZaberWiki
Jump to navigation Jump to search
Instruction Name Restore Settings
Applies to A-Series and X-Series Devices
Firmware Version 6.xx
Command Number 36
Command Type Command
Command Data 0 or Peripheral ID
Reply Data 0 or Peripheral ID
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Restores the device settings to the factory defaults.

This instruction is very useful for troubleshooting. If the device does not appear to function properly, it may be because some of the settings have been changed. This instruction will restore the settings to default values. All settings affected by this instruction are stored in non-volatile memory and will persist after power-down or reset.

A-Series and X-Series devices with an integrated controller will only accept a data value of 0.

Note: For A-Series and X-Series controllers, when used with a peripheral ID, this command is equivalent to sending Set Peripheral ID (Cmd 66) with a peripheral ID value. However, Set Peripheral ID with data 0 puts controller into safe-mode settings, whereas Restore Settings with data 0 causes the controller to load default values for the assigned peripheral ID.


Warning: This is an "include" page. As such, its contents may be included in multiple other pages. To see a list of all pages that include this page, click on "What links here" in the bottom left navbar. If you make changes to this page, be sure that your changes are appropriate to all pages on which this content will be included. You are also encouraged to review our Zaber Wiki Editing Guidelines. This header will not be included in other pages.