Manuals/A-BLQ-E

From ZaberWiki
< Manuals
Revision as of 13:20, 11 January 2012 by David Goosen (Talk | contribs)

Jump to: navigation, search
A-BLQ-E Belt-Driven Linear Stages
A-BLQ-E Belt-Driven Linear Stage



Contents



Disclaimer

Zaber’s devices are not intended for use in any critical medical, aviation, or military applications or situations where a product's use or failure could cause personal injury, death, or damage to property. Zaber disclaims any and all liability for injury or other damages resulting from the use of our products.

Precautions

Zaber's linear stages can produce enough force to cause personal injury. Be careful to not get body parts caught between the moving carriage and the end plates of the stage. Make sure that no loose clothing gets caught in the rotating lead screw.

Maintenance

Stages with teflon-coated (black) leadscrews are designed to be maintenance free. There is no need to grease the leadscrew periodically. If the leadscrew becomes dirty, it can be cleaned with isopropyl alcohol. The rails can also be cleaned with isopropyl alcohol. Stronger solvents are not recommended as they can damage non-metal components.

Initial Setup

If you are unfamiliar with Zaber motorized devices, you may wish to perform a few simple steps to familiarize yourself with their operation. First you will need a power supply. If you ordered one with your device, you shouldn’t have any problem. If not, you will require a power supply with output voltage between 24 and 48VDC. 48V will provide the fastest speed and the most power at high speeds. Power is not daisy chained through the data cables, so every A-Series device will require its own power supply.

Once you have a working power supply, you can test the operation of your device. Try turning the manual control knob counterclockwise to move the device towards its home position. The device is homed when it is first retracted completely. You must home the device every time you connect the power. You will not be able to achieve the maximum travel until you have homed the device. After the device retracts completely, it will stop and you can turn the knob clockwise to extend the device.

If you have not purchased a power supply with your Zaber device, you will have been shipped a PMF3 power plug to connect to your own supply. The wiring diagram below shows the power connector pinouts as seen looking into the connector on the device.


Pinout for Molex Mini-fit Jr. 3 pin (24-48V device power connector)

Molex mini-fit jr. 3pin
T-MCApowerconnector.jpg

Installation

Connection to a Computer

Connect the device to a computer as follows:

  1. Plug the Mini-DIN to D-Sub serial adaptor (T-DSUB9) into your computer’s serial port and plug the device’s data cable into the adaptor (figure 1). You may need to use a cable extension to reach your computer. There is no need to power-down or reboot the computer. USB-to-RS232 converters are available if you have no RS232 serial port.
  2. Connect the power plug of your power supply to the power connector of the device (figure 2). The green LED should light up indicating the device has power.
  3. Additional devices can simply be daisy-chained to the first (figure 3). A power supply should be connected to each device in the chain. A Renumber (Cmd 2) command needs to be issued after a new device is connected, so that the computer can control each device independently.
  4. Install software from http://www.zaber.com/wiki/Software (figure 4). We recommend installing the Zaber Console to get started. As a simple first test, try entering these instructions:
Renumber all devices                - Device: 0, Cmd:  2, Data: 0 
Home device 1                       - Device: 1, Cmd:  1, Data: 0
Move device 1 to 10,000 microsteps  - Device: 1, Cmd: 20, Data: 10000

To see the microstep distance (resolution) for your device, see the specs page for that device on the respective product page.


InstallationStep1.jpg InstallationStep2.jpg InstallationStep4.jpg InstallationStep3.jpg
Figure 1 Figure 2 Figure 3 Figure 4


Data Cable Wiring Diagram

Pin# Minidin 6 Female
(to next device)
Minidin 6 Male
(to previous device)
Minidin 6 female pinout.jpg Minidin 6 male pinout.jpg
1 - -
2 Receive (from next device) Transmit (toward computer)
3 Ground Ground
4 +12-16 Vdc * +12-16 Vdc *
5 - -
6 Transmit (to next device) Receive (from computer)
nc - -
* T-Series 12-16V devices only (all T-series devices, except T-LSQ, T-LST, T-MCA)

Note that multiple cable suppliers have been used over the years, each implementing different color codes. Rather than relying on the wire colors, it is recommended that users perform a continuity test to determine which wires are connected to which pins on their own device.

To connect a Zaber device to a computer serial port requires a T-DSUB9 adaptor. It looks similar to an old PS/2 serial mouse adaptor, but it has different wiring.


Physical Installation

Mounting

There are several options available for mounting Zaber stages. Use the mounting holes in the bottom to mount to a surface or to another stage. You might have to move the carriage to access the bottom mounting holes. Some stages have mounting holes in the end plates for mounting vertically. Mounting screws are included with most stages.

Caution: Some stages have threaded through-holes in the top mounting plate of the carriage. Be sure not to install mounting screws too deep, causing them to interfere with inside parts of the stage.

Grounding

To prevent damage to the device due to static buildup, the device should be properly grounded.

Failure to ground the unit may result in the unit shutting down unexpectedly or ceasing to communicate with the computer. This problem can be minimized by not touching the unit during operation. If the unit fails due to static discharge, unplugging it and plugging it back in or sending a Restore Settings command will usually fix the problem.

Most Zaber devices are grounded via the shield wire of the data cables. This should normally provide a path to ground via the computer. For units which are being used without a computer, a ground lead should be connected to the shield of one of the data cables.

Note: Encoder embedded devices are sensitive to electrically noisy environments. Static discharges can affect position calibration and cause unstable behaviour. If the device is behaving strangely, verify that the device is properly grounded, reset the device (either by the Reset (Cmd 0) command or by disconnecting then reconnecting power), then re-initialize the device with the Home (Cmd 1) command.


Tensioning The Timing Belt

Although A-BLQ stages ship with the timing belt properly tensioned, it may at some point become necessary to re-tension it. Maintaining the belt at its proper tension will ensure you are getting the best performance out of the stage. Included with each A-BLQ stage is a clip-on instrument tuner that, when used as described below, is very effective at setting the belt to a precise tension.


To tension the timing belt:

1. Plug stage into PC and move to absolute position of 68960. The tuning procedure works better if the stage is not bolted down to anything.
2. Attach instrument tuner to motor pulley housing as shown below. Tuner must not be touching the belt.
Position stage to absolute position of 68690 and attach tuner as shown.

3. Make sure the tuner is set to vibration ("VIB") mode. Turn on by pressing the large grey button on the front.
Make sure tuner is set to vibration mode.

4. Pluck the belt in between the motor pulley housing and the carriage. The tuner should register a note. When the belt is properly tensioned, this note should be C2 (65.4 Hz).
Pluck belt to measure tension.

Read the note picked up by the tuner.

5. Make adjustments to the belt tension with the M4 screw loosened about 1/4 turn from tight. The screw and the belt clamp under the stage top need to be able to slide along the slot. To increase the belt tension, turn the setscrew clockwise - pushing the belt clamp towards the middle of the stage top. To decrease belt tension, turn the setscrew counter-clockwise.
Loosen M4 screw then adjust setscrew to tighten or loosen belt.

6. Try to set the belt tension to about a B2 note. The tension will increase by about a semitone (e.g. B to C or C to C#) when the M4 screw is tightened. The final tension should be within 25 cents of C2 (1 semitone = 100 cents).
The final tension should be close to C2

7. The tuner may not initially register a note if the belt is too loose. As a rough guideline, you can hear what a C2 sounds like at www.tunemybass.com (select 4 string base with drop C tuning, then mouse over the lowest string to hear tone). It is a good idea to get a rough idea of what the tensioned belt will sound like so that you don't try to tune it to a C3 note - an octave higher than C2. This will be much too tight and could damage the belt or tensioning mechanism.
8. If the belt is close to the right tension, but the tuner still does not register a note, try attaching a metal plate or bracket to the stage top and clamp the tuner to that. This mounting should pick up the belt vibrations more easily.
If the tuner is having difficulty picking up the belt vibration, try indirectly attaching it to the stage top as shown.


Manual Control

Without connecting to a computer, Zaber's motion products can be manually controlled via a knob. A-Series products are integrated with a depressible knob with 20 detents per revolution.


There are two manual movement modes available. You can toggle between the two modes by holding down the knob for 1 sec or via Set Knob Movement Mode (Cmd 109).

Velocity Mode
  • Turn the knob clockwise to move the device in the positive direction (extend), counter-clockwise for negative direction (retract).
  • Each detent of the knob increases the speed of the carriage.
  • There are 16 speeds in each direction. The velocity profile and maximum speed can be configured via Set Knob Velocity Profile (Cmd 112) and Set Knob Velocity Scale (Cmd 111).
  • The device stops and resets the knob upon arriving at the end of travel.
Displacement Mode
  • Turn the knob clockwise to move the device in the positive direction (extend), counter-clockwise for negative direction (retract).
  • Each detent of the knob moves the device a fixed number of microsteps, specified by Set Knob Jog Size (Cmd 110).
  • The device moves at the speed specified by Set Target Speed (Cmd 42), or the slower of Target Speed and Home Speed if the device has not been homed.
  • The device stops at the end of travel if the number of microsteps left in the set range is less than the requested number.


Summary of knob functionality:

Turning the knob
Moves the device in the direction of knob turn
Pressing the knob
Decelerates and stops the device (identical to a Stop command)
Instantly stops the device, if the device is already decelerating
Pressing and holding the knob for 1 sec
Toggle between Velocity Mode and Displacement Mode


Requirements for using Manual Control without a Computer

After each power up, you must retract the device completely to home the device. You will not be able to extend the device past the power up position unless you have first retracted it completely to set the home position.

LED functions

A-Series Device LED Layout
G
R Y B


Green - Power
  • On: Device is operational
  • Flashing at 2Hz: Power supply voltage or temperature is out of range
  • Fading in and out slowly: Device is Parked. See Set Park State (Cmd 65)
Yellow - Communication/Busy
  • On: Device is moving, or traffic packets are being transferred on RS232 communication lines.
  • Blinking: Device is under manual control via the knob (in Velocity mode). Blinking rate is proportional to movement speed.
Blue - Slip/Stall
  • On: Device is slipping
  • 2 short flashes every 1 sec: The stationary device has been forced out of position
  • On-Off cycle every 2 sec: The device has stalled and stopped
Red - Error
  • On/flashing: Device has lost its settings, or an error has occurred. Please contact technical support.

Trajectory Control and Behaviour

This section describes the behaviour of the device trajectory when a movement command is issued.

Software Position Limits

The travel range of the device is confined by the Minimum Position and Maximum Position settings. The factory settings of the devices are configured to match the physical travel range. If a customized range is desired they can be changed via Set Minimum Position (Cmd 106) and Set Maximum Position (Cmd 44).


Minimum Position
When Current Position is less than the Minimum Position value, the device cannot move in the negative direction (towards the motor).
Maximum Position
When Current Position is greater than the Maximum Position value, the device cannot move in the positive direction (away from the motor).

Movement Speed

The movement speed of the device depends on device status and various speed settings. If the device has not been initialized by the Home (Cmd 1) command or moving towards the home end of the device, movement speed will be constrained to fail-safe values.

The home status of the device can be determined by sending the Return Setting (Cmd 53) command with a data value of 103 (command number for Set Home Status (Cmd 103)).


Movement speed of the device is specified below:

Move At Constant Speed (Cmd 22)
The device will move at the specified speed regardless of home status.
Knob manual movement in Velocity Mode
The device will move at the specified speed regardless of home status.
The speed is specified by Set Knob Velocity Profile (Cmd 112) and Set Knob Velocity Scale (Cmd 111).
Other movement commands - When device has not been homed
The device will move at the slower of Target Speed and Home Speed.
Other movement commands - When device has been homed
The device will move at the speed specified by Target Speed.

Control Through The RS-232 Serial Port

All Zaber devices use the same RS232 communications protocol. The yellow LED will light when there is activity on the RS232 lines. You may use this feature to determine which COM port you are connected to. We recommend using the Zaber Console that you can download from our web site. The source code is also available for you to use as an example for writing your own custom code. See the troubleshooting section later in this manual if you have trouble communicating with the device.

Serial Port Configuration - 9600/8-N-1
  • Baudrate: 9600 bits/s
  • Data bits: 8
  • Parity bit: None
  • Stop bit: 1
  • Handshake/Flow control: None

Important: The first time you connect a device to your computer you must issue a renumber instruction to assign each device a unique identifier. This should be done after all the devices in the daisy-chain are powered up. The device number is stored in non-volatile memory and will persist after powerdown, so you need only issue the renumber instruction when you add new devices to the chain, or rearrange the order of the devices, however it does no harm to issue the renumber instruction after every powerup. You must not transmit any instructions while the chain is renumbering or the renumbering routine may be corrupted. Renumbering takes less than a second, after which you may start issuing instructions over the RS232 connection.

All instructions consist of a group of 6 bytes. They must be transmitted with less than 10 ms between each byte. If the device has received less than 6 bytes and then a period longer than 10 ms passes, it ignores the bytes already received. We recommend that your software behave similarly when receiving data from the devices, especially in a noisy environment like a pulsed laser lab.

The following table shows the instruction format:

  • Byte 1 - Device #
  • Byte 2 - Command #
  • Byte 3 - Data - Least Significant Byte (LSB)
  • Byte 4 - Data
  • Byte 5 - Data
  • Byte 6 - Data - Most Significant Byte (MSB)

The first byte is the device number in the daisy-chain. Device number 1 is the closest device to the computer and device number 2 is next and so on. If the number 0 is used, all the devices in the chain will process the accompanying command simultaneously.

The second byte is the command number. Bytes 3, 4, 5, and 6 are data in long integer, 2’s complement format with the least significant byte transmitted first. How the command data are interpreted depends on the command. Complete details are given in the command reference on the following page.

Examples

  • All devices renumber: 0, 2, 0, 0, 0, 0
  • All devices home: 0, 1, 0, 0, 0, 0
  • All devices return firmware version: 0, 51, 0, 0, 0, 0
  • Device 1 move to an absolute position (command 20) of 257 microsteps: 1, 20, 1, 1, 0, 0
  • Device 2 move to a relative position (command 21) of -1 microstep: 2, 21, 255, 255, 255, 255

If you are using Zaber’s demo software, you will only see 3 entry fields: Device#, Command#, and Data. The Device# and Command# fields accept unsigned integer values while the value you enter into the Data field can be signed. The value in the data field is converted by the software to 4 separate bytes and then gets sent to the device.

Most instructions cause the device to reply with a return code. It is also a group of 6 bytes. The first byte is the device #. Byte #2 is the instruction just completed or 255 (0xFF) if an error occurs. Bytes 3, 4, 5 and 6 are data bytes in the same format as the instruction command data.

Data Conversion Algorithms

If you are writing software to control Zaber products, you'll likely need to generate data bytes 3 through 6 from a single data value, or vise versa. The following pseudo-code can be used as a model.

Converting command data into command bytes to send to Zaber products

If Cmd_Data < 0 then Cmd_Data = 256^4 + Cmd_Data                                                'Handles negative data
Cmd_Byte_6 = Cmd_Data / 256^3
Cmd_Data   = Cmd_Data - 256^3 * Cmd_Byte_6
Cmd_Byte_5 = Cmd_Data / 256^2
Cmd_Data   = Cmd_Data - 256^2 * Cmd_Byte_5
Cmd_Byte_4 = Cmd_Data / 256
Cmd_Data   = Cmd_Data - 256   * Cmd_Byte_4
Cmd_Byte 3 = Cmd_Data

Converting reply bytes into a single reply data value

Reply_Data = 256^3 * Rpl_Byte 6 + 256^2 * Rpl_Byte_5 + 256 * Rpl_Byte_4 + Rpl_Byte_3
If Rpl_Byte_6 > 127 then Reply_Data = Reply_Data - 256^4                                        'Handles negative data


Sample Waveforms

If you are designing hardware to interface with Zaber products, it may be useful to see some sample waveforms.

This image shows an instruction (top waveform) and a reply (bottom waveform). Rs232 waveform instruction and reply.jpg
This is a closeup of the instruction bytes. Note that the voltage swing of the instruction waveform is about +/-10V. This is typical of the USB to RS232 converter to which the device is connected. The instruction bytes are 0, 51, 0, 0, 0, 0 indicating the instruction: Device: All, Command: Return version, Data: 0. Rs232 waveform instruction.jpg
This is a closeup of the instruction byte 2. The "S" on either end represents the start and stop bits respectively. The start bit is always positive and the stop bit is always negative. The bit sequence is Start, 0, 1, 2, 3, 4, 5, 6, 7, Stop. The data is output with the least significant bit first, so the byte shown is actually 00110011 in binary or 32+16+2+1 = 51 in decimal.

It's a good idea to do a timing "reality check" whenever viewing a waveform such as this. The bit rate should be 9600 bits/second. The time scale in the image is 200 us/division. Since a byte consists of 10 bits (start, stop, plus 8 data bits) and the bit rate is 9600 bits/s, the duration of a single byte should be 1.04 ms (10/9600 seconds) or about 5.2 divisions at 200 us/division. It can be seen in the image that this is the case.

Rs232 waveform instruction byte.jpg
This is a closeup of the reply bytes. Note that the voltage swing of the reply waveform is about +/-8V. This is a typical output from the RS232 driver chip used in Zaber devices. The instruction was to all devices. Only a single device is present so there is only one reply. The reply is 1, 51, 252, 1, 0, 0 indicating the reply "Device: 1, Reply: Firmware version, Data: 508". The data is the last 4 bytes combined, with least significant byte first. In this case the data is 0*256^3 + 0*256^2 + 1*256 + 252 = 508. This indicates a firmware version of 5.08 since according to the instruction specification a decimal is assumed before the last two digits.

The time scale in this image is 1 ms/division. As noted above, at a bit rate of 9600 bits/second each byte should take 1.04 ms. Therefore a 6 byte reply should take 6.24 ms. It can be seen in the image that this is the case.

Rs232 waveform reply.jpg


Quick Command Reference

The following table offers a quick command reference for motorized closed-loop devices running firmware version 6.xx. For convenience, you may sort the table below by instruction name, command number, or reply number. Follow the links to view a detailed description of each instruction.

Instruction Name Command# Command Data Command Type Reply Data
Reset 0 Ignored Command None
Home 1 Ignored Command Final position (in this case 0)
Renumber* 2 Ignored Command Device Id
Read Register 5 Register Address Command Data
Set Active Register 6 Register Address Setting Register Address
Write Register 7 Data Command Data
Move Tracking 8 n/a Reply Tracking Position
Limit Active 9 n/a Reply Final Position
Manual Move Tracking 10 n/a Reply Tracking Position
Manual Move 11 n/a Reply Final Position
Slip Tracking 12 n/a Reply Tracking Position
Unexpected Position 13 n/a Reply Final Position
Store Current Position* 16 Address Command Address
Return Stored Position 17 Address Command Stored Position
Move To Stored Position 18 Address Command Final Position
Move Absolute 20 Absolute Position Command Final Position
Move Relative 21 Relative Position Command Final Position
Move At Constant Speed 22 Speed Command Speed
Stop 23 Ignored Command Final Position
Restore Settings* 36 Peripheral Id Command Peripheral Id
Set Microstep Resolution* 37 Microsteps Setting Microsteps
Set Running Current* 38 Value Setting Value
Set Hold Current* 39 Value Setting Value
Set Device Mode* 40 Mode Setting Mode
Set Target Speed* 42 Speed Setting Speed
Set Acceleration* 43 Acceleration Setting Acceleration
Set Maximum Position* 44 Range Setting Range
Set Current Position 45 New Position Setting New Position
Set Home Offset* 47 Offset Setting Offset
Set Alias Number* 48 Alias Number Setting Alias Number
Return Device Id 50 Ignored Read-Only Setting Device Id
Return Firmware Version 51 Ignored Read-Only Setting Version
Return Power Supply Voltage 52 Ignored Read-Only Setting Voltage
Return Setting 53 Setting Number Command Setting Value
Return Status 54 Ignored Read-Only Setting Status
Echo Data 55 Data Command Data
Return Current Position 60 Ignored Read-Only Setting Position
Set Park State* 65 Park State Setting Position
Set Auto-Reply Disabled Mode* 101 Auto-Reply Mode Setting Auto-Reply Mode
Set Message Id Mode* 102 Message Id Mode Setting Message Id Mode
Set Home Status 103 Home Status Setting Home Status
Set Minimum Position* 106 Minimum Position Setting Minimum Position
Set Knob Disabled Mode* 107 Knob Disabled Mode Setting Knob Disabled Mode
Set Knob Direction* 108 Direction Setting Direction
Set Knob Movement Mode* 109 Movement Mode Setting Movement Mode
Set Knob Jog Size* 110 Jog Size Setting Jog Size
Set Knob Velocity Scale* 111 Velocity Scale Setting Velocity Scale
Set Knob Velocity Profile* 112 Velocity Profile Setting Velocity Profile
Set Acceleration Only* 113 Acceleration Setting Acceleration
Set Deceleration Only* 114 Deceleration Setting Deceleration
Set Move Tracking Mode* 115 Tracking Mode Setting Tracking Mode
Set Manual Move Tracking Disabled Mode* 116 Tracking Mode Setting Tracking Mode
Set Move Tracking Period* 117 Tracking Period Setting Tracking Period
Set Closed-Loop Mode* 118 Closed-Loop Mode Setting Closed-Loop Mode
Set Slip Tracking Period* 119 Tracking Period Setting Tracking Period
Set Stall Timeout* 120 Timeout Setting Timeout
Error 255 n/a Reply Error Code

* The settings for these commands are saved in non-volatile memory, i.e. the setting persists even if the device is powered down. To restore all settings to factory default, use command 36.

This manual describes the binary communications protocol as supported by all of Zaber's A-Series devices. For devices with firmware version 6.06 and above, the ASCII Protocol is the recommended method of communication.

Quick Start

  1. Download Zaber Console from http://www.zaber.com/wiki/Software/Zaber_Console and install it.
  2. Switch to the Options Tab and check that the Baud Rate is set to 9600 and the Protocol to Binary.
    ZaberConsole Options.png
  3. Select the Serial Port to use and Click the Open button. A list of attached devices will be shown.
    ZaberConsole Devices.png
  4. Switch to the Commands tab:
    1. Enter 0 in the box next to Renumber and press Send.
    2. Press the Send button next to the home command.
    3. Enter 10000 in the box next to Move Absolute and press send.
      ZaberConsole Commands.png

Serial Port Configuration

All devices using the binary protocol default to the following settings:

  • Baudrate: 9600 bits/s
  • Data bits: 8
  • Parity bit: None
  • Stop bit: 1
  • Handshake/Flow control: None

Message Format

All instructions consist of a group of 6 bytes. They must be transmitted with less than 10 ms between each byte. If the device has received less than 6 bytes and then a period longer than 10 ms passes, it ignores the bytes already received. We recommend that your software behave similarly when receiving data from the devices, especially in a noisy environment like a pulsed laser lab.

The following table shows the instruction format:

  • Byte 1 - Device #
  • Byte 2 - Command #
  • Byte 3 - Data - Least Significant Byte (LSB)
  • Byte 4 - Data
  • Byte 5 - Data
  • Byte 6 - Data - Most Significant Byte (MSB)

The first byte is the device number in the daisy-chain. Device number 1 is the closest device to the computer and device number 2 is next and so on. If the number 0 is used, all the devices in the chain will process the accompanying command simultaneously.

The second byte is the command number. Bytes 3, 4, 5, and 6 are data in long integer, 2’s complement format with the least significant byte transmitted first. How the command data are interpreted depends on the command. Complete details are given in the command reference on the following page.

Examples

  • All devices renumber: 0, 2, 0, 0, 0, 0
  • All devices home: 0, 1, 0, 0, 0, 0
  • All devices return firmware version: 0, 51, 0, 0, 0, 0
  • Device 1 move to an absolute position (command 20) of 257 microsteps: 1, 20, 1, 1, 0, 0
  • Device 2 move to a relative position (command 21) of -1 microstep: 2, 21, 255, 255, 255, 255

If you are using Zaber’s demo software, you will only see 3 entry fields: Device#, Command#, and Data. The Device# and Command# fields accept unsigned integer values while the value you enter into the Data field can be signed. The value in the data field is converted by the software to 4 separate bytes and then gets sent to the device.

Most instructions cause the device to reply with a return code. It is also a group of 6 bytes. The first byte is the device #. Byte #2 is the instruction just completed or 255 (0xFF) if an error occurs. Bytes 3, 4, 5 and 6 are data bytes in the same format as the instruction command data.

Data Conversion Algorithms

If you are writing software to control Zaber products, you'll likely need to generate data bytes 3 through 6 from a single data value, or vise versa. The following pseudo-code can be used as a model.

Converting command data into command bytes to send to Zaber products

If Cmd_Data < 0 then Cmd_Data = 256^4 + Cmd_Data                                                'Handles negative data
Cmd_Byte_6 = Cmd_Data / 256^3
Cmd_Data   = Cmd_Data - 256^3 * Cmd_Byte_6
Cmd_Byte_5 = Cmd_Data / 256^2
Cmd_Data   = Cmd_Data - 256^2 * Cmd_Byte_5
Cmd_Byte_4 = Cmd_Data / 256
Cmd_Data   = Cmd_Data - 256   * Cmd_Byte_4
Cmd_Byte 3 = Cmd_Data

Converting reply bytes into a single reply data value

Reply_Data = 256^3 * Rpl_Byte 6 + 256^2 * Rpl_Byte_5 + 256 * Rpl_Byte_4 + Rpl_Byte_3
If Rpl_Byte_6 > 127 then Reply_Data = Reply_Data - 256^4                                        'Handles negative data


Sample Waveforms

If you are designing hardware to interface with Zaber products, it may be useful to see some sample waveforms.

This image shows an instruction (top waveform) and a reply (bottom waveform). Rs232 waveform instruction and reply.jpg
This is a closeup of the instruction bytes. Note that the voltage swing of the instruction waveform is about +/-10V. This is typical of the USB to RS232 converter to which the device is connected. The instruction bytes are 0, 51, 0, 0, 0, 0 indicating the instruction: Device: All, Command: Return version, Data: 0. Rs232 waveform instruction.jpg
This is a closeup of the instruction byte 2. The "S" on either end represents the start and stop bits respectively. The start bit is always positive and the stop bit is always negative. The bit sequence is Start, 0, 1, 2, 3, 4, 5, 6, 7, Stop. The data is output with the least significant bit first, so the byte shown is actually 00110011 in binary or 32+16+2+1 = 51 in decimal.

It's a good idea to do a timing "reality check" whenever viewing a waveform such as this. The bit rate should be 9600 bits/second. The time scale in the image is 200 us/division. Since a byte consists of 10 bits (start, stop, plus 8 data bits) and the bit rate is 9600 bits/s, the duration of a single byte should be 1.04 ms (10/9600 seconds) or about 5.2 divisions at 200 us/division. It can be seen in the image that this is the case.

Rs232 waveform instruction byte.jpg
This is a closeup of the reply bytes. Note that the voltage swing of the reply waveform is about +/-8V. This is a typical output from the RS232 driver chip used in Zaber devices. The instruction was to all devices. Only a single device is present so there is only one reply. The reply is 1, 51, 252, 1, 0, 0 indicating the reply "Device: 1, Reply: Firmware version, Data: 508". The data is the last 4 bytes combined, with least significant byte first. In this case the data is 0*256^3 + 0*256^2 + 1*256 + 252 = 508. This indicates a firmware version of 5.08 since according to the instruction specification a decimal is assumed before the last two digits.

The time scale in this image is 1 ms/division. As noted above, at a bit rate of 9600 bits/second each byte should take 1.04 ms. Therefore a 6 byte reply should take 6.24 ms. It can be seen in the image that this is the case.

Rs232 waveform reply.jpg



Quick Command Reference

The following table offers a quick command reference for devices running firmware version 6.00 and above. For convenience, you may sort the table below by instruction name, command number, or reply number. Follow the links to view a detailed description of each instruction.

Instruction Name Command# Command Data Command Type Reply Data
Reset 0 Ignored Command None
Home 1 Ignored Command Final position (in this case 0)
Renumber* 2 Ignored Command Device Id
Read Register 5 Register Address Command Data
Set Active Register 6 Register Address Setting Register Address
Write Register 7 Data Command Data
Move Tracking 8 n/a Reply Tracking Position
Limit Active 9 n/a Reply Final Position
Manual Move Tracking 10 n/a Reply Tracking Position
Manual Move 11 n/a Reply Final Position
Slip Tracking 12 n/a Reply Tracking Position
Unexpected Position 13 n/a Reply Final Position
Store Current Position* 16 Address Command Address
Return Stored Position 17 Address Command Stored Position
Move To Stored Position 18 Address Command Final Position
Move Absolute 20 Absolute Position Command Final Position
Move Relative 21 Relative Position Command Final Position
Move At Constant Speed 22 Speed Command Speed
Stop 23 Ignored Command Final Position
Restore Settings* 36 Peripheral Id Command Peripheral Id
Set Microstep Resolution* 37 Microsteps Setting Microsteps
Set Running Current* 38 Value Setting Value
Set Hold Current* 39 Value Setting Value
Set Device Mode* 40 Mode Setting Mode
Set Target Speed* 42 Speed Setting Speed
Set Acceleration* 43 Acceleration Setting Acceleration
Set Maximum Position* 44 Range Setting Range
Set Current Position 45 New Position Setting New Position
Set Home Offset* 47 Offset Setting Offset
Set Alias Number* 48 Alias Number Setting Alias Number
Return Device Id 50 Ignored Read-Only Setting Device Id
Return Firmware Version 51 Ignored Read-Only Setting Version
Return Power Supply Voltage 52 Ignored Read-Only Setting Voltage
Return Setting 53 Setting Number Command Setting Value
Return Status 54 Ignored Read-Only Setting Status
Echo Data 55 Data Command Data
Set Peripheral Id* 66 Peripheral Id Setting Peripheral Id
Return Current Position 60 Ignored Read-Only Setting Position
Set Park State* 65 Park State Setting Position
Set Auto-Reply Disabled Mode* 101 Auto-Reply Mode Setting Auto-Reply Mode
Set Message Id Mode* 102 Message Id Mode Setting Message Id Mode
Set Home Status 103 Home Status Setting Home Status
Set Home Sensor Type* 104 Home Sensor Type Setting Home Sensor Type
Set Auto-Home Disabled Mode* 105 Auto-Home Disabled Mode Setting Auto-Home Disabled Mode
Set Minimum Position* 106 Minimum Position Setting Minimum Position
Set Knob Disabled Mode* 107 Knob Disabled Mode Setting Knob Disabled Mode
Set Knob Direction* 108 Direction Setting Direction
Set Knob Movement Mode* 109 Movement Mode Setting Movement Mode
Set Knob Jog Size* 110 Jog Size Setting Jog Size
Set Knob Velocity Scale* 111 Velocity Scale Setting Velocity Scale
Set Knob Velocity Profile* 112 Velocity Profile Setting Velocity Profile
Set Acceleration Only* 113 Acceleration Setting Acceleration
Set Deceleration Only* 114 Deceleration Setting Deceleration
Set Move Tracking Mode* 115 Tracking Mode Setting Tracking Mode
Set Manual Move Tracking Disabled Mode* 116 Tracking Mode Setting Tracking Mode
Set Move Tracking Period* 117 Tracking Period Setting Tracking Period
Set Closed-Loop Mode* 118 Closed-Loop Mode Setting Closed-Loop Mode
Set Slip Tracking Period* 119 Tracking Period Setting Tracking Period
Set Stall Timeout* 120 Timeout Setting Timeout
Set Device Direction* 121 Direction Setting Direction
Set Baudrate* 122 Baudrate Setting Baudrate
Set Protocol* 123 Protocol Setting Protocol
Convert To Ascii* 124 Baudrate Setting Baudrate
Error 255 n/a Reply Error Code

* The settings for these commands are saved in non-volatile memory, i.e. the setting persists even if the device is powered down. To restore all settings to factory default, use command 36.

Command Reference

Reset - Cmd 0

Instruction Name Reset
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 0
Command Type Command
Command Data Ignored
Reply Data None
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Sets the device to its power-up condition.

This has the same effect as unplugging and restarting the device.

Special Note

The position of the device will be lost upon reset. If it is important to keep the same position, Park the stage before resetting/powering down and Unpark the stage afterwards to recover the position settings (See Set Park State (Cmd 65) command).

All non-volatile settings (i.e. Device Number, Target Velocity, etc.) are saved and are not affected by reset or power-down.

Home - Cmd 1

Instruction Name Home
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 1
Command Type Command
Command Data Ignored
Reply Data Final Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Moves to the home position and resets the device's internal position.

Upon receiving this instruction, the device will retract until its internal home sensor is triggered. It will then move forward several steps to avoid accidentally re-triggering the home sensor during use. Its internal position is then reset (to 0 for most devices). If a home offset has been specified with the Set Home Offset (cmd 47) instruction, the device will move forward for the specific offset, then reset the internal position.

The device will attempt to home for an extended amount of time. If Closed-Loop Mode is enabled for an encoder-embedded device, the device will report any slip or stall incident.

NOTE: The device will attempt to move towards the home sensor regardless of the Minimum Position and Home Offset settings. Set Minimum Position (Cmd 106) does not prevent the device from moving pass the Minimum Position during Homing. If you need to initialize the device without homing it, Park the stage before power down and Unpark the stage when power is restored. See Set Park State (Cmd 65) for details.

Renumber - Cmd 2

Instruction Name Renumber
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 2
Command Type Command
Command Data New Number
Reply Data Device Id
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Assigns new numbers to all the devices in the order in which they are connected.

This command is usually sent to device number 0. When it is, the command data is ignored and all devices will renumber. The device closest to the computer becomes device number 1. The next device becomes number 2 and so on.

If sent to a device number other than 0, then that device will reassign itself the device number in the command data. Valid device numbers are 1-99 for version 6.05, and 1-254 otherwise.

Note: Renumbering takes about 1/2 a second during which time the computer must not send any further data. The device number is stored in non-volatile memory so you can renumber once and not worry about issuing the renumber instruction again after each power-up.

Read Register - Cmd 5

Instruction Name Read Register
Applies to A-Series devices
Firmware Version 6.xx
Command Number 5
Command Type Command
Command Data Register Address
Reply Data Data
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Reads from a register.

This command reads the data value from a register location.

NOTE: This command is meant for Zaber's internal use. Do not use this command unless directed to by Zaber technical support.

Set Active Register - Cmd 6

Instruction Name Set Active Register
Applies to A-Series devices
Firmware Version 6.xx
Command Number 6
Command Type Setting
Command Data Register Address
Reply Data Register Address
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Sets the active register.

This command sets the address of the active register. When a device receives a Write Register (Cmd 7) command, data is written to the active register.

NOTE: This command is meant for Zaber's internal use. Do not use this command unless directed to by Zaber technical support.

Write Register - Cmd 7

Instruction Name Write Register
Applies to A-Series devices
Firmware Version 6.xx
Command Number 7
Command Type Command
Command Data Data
Reply Data Data
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Writes to a register.

This command writes the data to the active register. The active register is specified by Set Active Register (Cmd 6).

NOTE: This command is meant for Zaber's internal use. Do not use this command unless directed to by Zaber technical support.

Store Current Position - Cmd 16

Instruction Name Store Current Position
Applies to All motorized devices
Firmware Version 5.04 and up
Command Number 16
Command Type Command
Command Data Address
Reply Data Address
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Saves the current absolute position of the device.

Valid Address values are 0 through 15 specifying one of 16 possible registers in which to store the position. This command can only be executed when the device has been homed. This command is used in conjunction with the Return Stored Position (Command #17) and Move To Stored Position (Command #18) instructions. The positions stored in the position registers are non-volatile and will persist after power-down or reset. All position registers are cleared by the Restore Settings (Command #36) instruction.

Return Stored Position - Cmd 17

Instruction Name Return Stored Position
Applies to All motorized devices
Firmware Version 5.04 and up
Command Number 17
Command Type Command
Command Data Address
Reply Data Stored Position
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the position stored in one of the 16 position registers for the device.

Valid command data values are 0 through 15, specifying one of 16 possible registers from which to retrieve the position. This command is used in conjunction with the Store Current Position (#16) and Move To Stored Position (#18) commands. Positions stored in the position registers are non-volatile and will persist after power-down or reset. All position registers are cleared by the Restore Settings (#36) command.

Move To Stored Position - Cmd 18

Instruction Name Move To Stored Position
Applies to All motorized devices
Firmware Version 5.04 and up
Command Number 18
Command Type Command
Command Data Address
Reply Data Final Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Moves the device to the stored position specified by the Command Data.

Valid address values are 0 through 15, specifying one of 16 possible positions. This command is used in conjunction with the Store Current Position (#16) and Return Stored Position (#17) commands. This command does not send a response until the move has finished. All move commands are pre-emptive. If a new move command is issued before the previous move command is finished, the device will immediately move to the new position.

The target speed and acceleration during a move absolute instruction can be specified using Set Target Speed (Cmd 42) and Set Acceleration (Cmd 43) respectively.

This command may pre-empt, or be pre-empted by Move to Stored Position (Cmd 18), Move Absolute (Cmd 20), Move Relative (Cmd 21), Move at Constant Speed (Cmd 22) and Stop (Cmd 23).

Move Absolute - Cmd 20

Instruction Name Move Absolute
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 20
Command Type Command
Command Data Absolute Position
Reply Data Final position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Moves the device to the position specified in the Command Data in microsteps.

The device begins to move immediately, and sends a response when the move has finished. The position must be between Minimum and Maximum Position (specified by Set Minimum Position (Cmd 106) and Set Maximum Position (cmd 44)), or an error code will be returned.

The target speed and acceleration during a move absolute instruction can be specified using Set Target Speed (Cmd 42), Set Acceleration (Cmd 43), Set Acceleration Only (Cmd 113) and Set Deceleration Only (Cmd 114).

All move commands are pre-emptive. If a new move command is issued before the previous move command is finished, the device will immediately move to the new position. This command may pre-empt, or be pre-empted by Move to Stored Position (Cmd 18), Move Absolute (Cmd 20), Move Relative (Cmd 21), Move at Constant Speed (Cmd 22) and Stop (Cmd 23).

Move Relative - Cmd 21

Instruction Name Move Relative
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 21
Command Type Command
Command Data Relative Position
Reply Data Final Position
Safe to retry? No
Returns Current Position? Yes
Persistence n/a
Summary Moves the device by the positive or negative number of microsteps specified in the Command Data.

The device moves to a position given by its current position plus the value specified in the command data. The relative move command data in microsteps can be positive or negative. The final position must be between Minimum and Maximum Position (specified by Set Minimum Position (Cmd 106) and Set Maximum Position (cmd 44)), or an error code will be returned. The device begins to move immediately, and sends a response when the move has finished.

The target speed and acceleration during a move absolute instruction can be specified using Set Target Speed (Cmd 42), Set Acceleration (Cmd 43), Set Acceleration Only (Cmd 113) and Set Deceleration Only (Cmd 114).

All move commands are pre-emptive. If a new move command is issued before the previous move command is finished, the device will immediately move to the new position. If a Move Relative command is issued while the device is currently moving due to a previous command, the device will immediately set a new target position equal to the current position (at the instant the command was received) plus the specified relative position.

This command may pre-empt, or be pre-empted by Move to Stored Position (Cmd 18), Move Absolute (Cmd 20), Move Relative (Cmd 21), Move at Constant Speed (Cmd 22) and Stop (Cmd 23).

Move At Constant Speed - Cmd 22

Instruction Name Move At Constant Speed
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 22
Command Type Command
Command Data Speed
Reply Data Speed
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Moves the device at a constant speed based on the value specified in the Command Data.

This instruction specifies a direction and a speed to move, rather than a target position. When this instruction is issued the device will accelerate to the speed specified by the instruction data. The device will continue moving at this speed until a limit is reached or a pre-empting instruction is issued. Negative speeds cause retraction while positive speeds cause extension. Unlike the other movement commands, this command sends a response immediately without waiting for the move to finish.

If the device reaches minimum or maximum position (specified by Set Minimum Position (Cmd 106) and Set Maximum Position (cmd 44)), the device stops and the new position is returned via reply-only command Limit Active (Cmd 9).

The device accelerates/decelerates at a rate specified by Set Acceleration (Cmd 43), Set Acceleration Only (Cmd 113) and Set Deceleration Only (Cmd 114).

The device may be set to return its position continually during the move using Set Move Tracking Mode (Cmd 115).

This command may pre-empt, or be pre-empted by Move to Stored Position (Cmd 18), Move Absolute (Cmd 20), Move Relative (Cmd 21), Move at Constant Speed (Cmd 22) and Stop (Cmd 23).

For a spreadsheet that can be used to calculate speed setting values for any product see http://www.zaber.com/documents/ZaberSpeedSetting.xls. Alternatively you may use the formulas below.

Actual Speed

= Data / 1.6384 * M mm/s or deg/s
= Data / 1.6384 microsteps/s
= Data / 1.6384 / R steps/s
= Data / 1.6384 / (R x S) * 60 revolutions/min Motor rpm
= Data / 1.6384 * L / (R x S) mm/s Linear devices only

where:

  • Data is the value of the command data
  • R (microsteps/step) is the microstep resolution (command 37)
  • S (steps/revolution) is the number of steps per motor revolution
  • M (mm or deg) is the microstep size
  • L (mm or deg) is the distance of linear motion per motor revolution

Valid data values are from 1 to (16384*R). Note that the maximum speed possible is independent of the resolution.

Refer to product specifications for the distance corresponding to a single microstep or revolution.

For example, if a motor has 200 steps per revolution (S = 200), used with default resolution (R = 64), and Data is 153600, then the motor will move at a speed of 720 revolutions per minute.

Stop - Cmd 23

Instruction Name Stop
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 23
Command Type Command
Command Data Ignored
Reply Data Final Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Stops the device from moving by preempting any move instruction.

This instruction can be used to pre-empt any move instruction. The device will decelerate to a stop. The reply data is the absolute position after stopping.

Rate of deceleration can be specified by either Set Acceleration (Cmd 43) (sets both Acceleration and Deceleration) or Set Deceleration Only (Cmd 114) (sets Deceleration only).

Depressing the manual control knob also gives the same effect as this command.

If another Stop is issued during deceleration (either by this command or depressing the manual control knob), the device will instantly stop.

This command may pre-empt, or be pre-empted by Move to Stored Position (Cmd 18), Move Absolute (Cmd 20), Move Relative (Cmd 21), Move at Constant Speed (Cmd 22) and Stop (Cmd 23).

Restore Settings - Cmd 36

Instruction Name Restore Settings
Applies to A-Series Devices
Firmware Version 6.xx
Command Number 36
Command Type Command
Command Data 0 or Peripheral Id
Reply Data 0 or Peripheral Id
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Restores the device settings to the factory defaults.

This instruction is very useful for troubleshooting. If the device does not appear to function properly, it may be because some of the settings have been changed. This instruction will restore the settings to default values. All settings affected by this instruction are stored in non-volatile memory and will persist after power-down or reset.

A-Series devices with an integrated controller will only accept a data value of 0.

Note: For A-Series controllers, when used with a Peripheral Id, this command is equivalent to sending Set Peripheral Id (Cmd 66) with a peripheral Id value. However, Set Peripheral Id with data 0 puts controller into safe-mode settings, whereas Restore Settings with data 0 causes the controller to load default values for the assigned peripheral Id.

Set Microstep Resolution - Cmd 37

Instruction Name Set Microstep Resolution
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 37
Command Type Setting
Command Data Microsteps
Reply Data Microsteps
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes the number of microsteps per step.

This command sets the microstep resolution of a device.

This setting is stored in non-volatile memory and will persist after power-down or reset. Use Restore Settings (Cmd 36) to restore all non-volatile settings to factory default.

The default value on most devices is 64. Available microstep resolutions are:

1, 2, 4, 8, 16, 32, 64, 128, 256
3, 6, 12, 24, 48, 96, 192
5, 10, 20, 40, 80, 160
9, 18, 36, 72, 144
15, 30, 60, 120, 240
25, 50, 100, 200
27, 54, 108, 216
45, 90, 180

All position related settings are in units of microsteps. When you change the microstep resolution, microstep related settings are automatically restored to their default values and then adjusted to match the scale of the new resolution.

If you have previously changed any settings, you will need to reconfigure them after setting a new microstep resolution. The device should be homed after assigning a new microstep resolution.

The following example shows affected settings when resolution is changed from 64 to 32:

Setting Default Before After
Microstep Resolution * 64 64 32
Current Position n/a 10501 ** 5250 **
Target Speed * 153600 ignored 76800
Knob Velocity Scale * 153600 ignored 76800
Home Speed * 50000 ignored 25000
Maximum Position * 280000 ignored 140000
Minimum Position * 0 ignored 0
Home Offset * 0 ignored 0
Acceleration * 205 ** ignored 102 **
Deceleration * 205 ** ignored 102 **

* The settings for these commands are saved in non-volatile memory.

** Note that if a number is divided by two, it is rounded down to the nearest whole number.


Set Running Current - Cmd 38

Instruction Name Set Running Current
Applies to T-LSQ, T-LST, T-MCA, A-series motorized devices
Firmware Version 5.00 and up
Command Number 38
Command Type Setting
Command Data Value
Reply Data Value
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the desired current to be used when the device is moving.

If your application does not require high torque, it is best to decrease the driving current to reduce power consumption, vibration, and motor heating. Trial and error should suggest an appropriate setting. If higher torque is required, it is generally safe to overdrive motors as long as they are not operated continuously. Motor temperature is typically the best indication of the degree to which overdriving can be employed. If the motor gets too hot to touch (>75°C), you should reduce the running current.

The current is related to the data by the formula:

Current = CurrentCapacity * CommandData / 100

The Command Data can be a value from 0 (no current) to 100 (max). CurrentCapacity is the hardware's maximum capability of output current.

To prevent damage, some devices limit the maximum output current to a lower value. In that case the valid range is 0 - Limit. Current limits are listed under the device specifications.

Some devices limit the voltage rather than the current. In this case the same formula can be used by replacing Current and CurrentCapacity with Voltage and PowerSupplyVoltage.


For example, the T-MCA has a CurrentCapacity of 2.5A. To apply a current of 0.5A to a motor, reverse the formula:

CommandData

= Current / CurrentCapacity * 100
= 0.5A / 2.5A * 100
= 20

Therefore CommandData = 20.

Set Hold Current - Cmd 39

Instruction Name Set Hold Current
Applies to T-LSQ, T-LST, T-MCA, A-series motorized devices
Firmware Version 5.00 and up
Command Number 39
Command Type Setting
Command Data Value
Reply Data Value
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the desired current to be used when the device is holding its position.

It is typical to run stepper motors at their rated current only during moves (for highest torque) and reduce the current when idle just to hold the position.

Typically the hold current can be set to around 25 - 50% of the running current. In some applications, the friction of the drive system alone is sufficient to hold the microstep position of the motor, and the hold current can be turned off completely. The hold current can be turned off by issuing the "Set Hold Current" instruction with data of 0.

When the device is moving, it applies running current to the motor. When the device stops moving, running current is applied for an additional 0.1 second before hold current is applied.

The current is related to the data by the formula:

Current = CurrentCapacity * CommandData / 100

The Command Data can be a value from 0 (no current) to 100 (max). CurrentCapacity is the hardware's maximum capability of output current.

To prevent damage, some devices limit the maximum output current to a lower value. In that case the valid range is 0 - Limit. Current limits are listed under the device specifications.

Some devices limit the voltage rather than the current. In this case the same formula can be used by replacing Current and CurrentCapacity with Voltage and PowerSupplyVoltage.


For example, the T-MCA has a CurrentCapacity of 2.5A. To apply a current of 0.5A to a motor, reverse the formula:

CommandData

= Current / CurrentCapacity * 100
= 0.5A / 2.5A * 100
= 20

Therefore CommandData = 20.

Set Device Mode - Cmd 40

Instruction Name Set Device Mode
Applies to A-Series motor controllers
Firmware Version 6.xx
Command Number 40
Command Type Setting
Command Data Mode
Reply Data Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the Device Mode for the given device.

This command allows setting several options, each controlled by a single bit within the command data. In A-Series devices, this command is replaced by individual setting commands. This command is deprecated but is supported for backward compatibility.

To determine what decimal value to use, the command data may be considered as a single 32-bit binary value. The least significant bit is bit_0, the next is bit_1, the next is bit_2, and so on up to the most significant bit_31. Each bit may have a value of either 1 or 0. Any unused or reserved bits should be left as 0.

The corresponding decimal representation of this 32-bit data is given by:

Decimal value = (bit_0 * 1) + (bit_1 * 2) + … + (bit_n * 2^n) + … + (bit_31 * 2^31)

For example, suppose you want all mode bits to be 0 except for bit_3 (disable knob) and bit_6 (enable message Ids). The Set Device Mode instruction should be sent with data calculated as follows:

Command Data
= 2^3 + 2^6
= 8 + 64
= 72

Note that each instance of the Set Device Mode command overwrites ALL previous mode bits. Repeated commands do not have a cumulative effect. For example, suppose you send a Set Device Mode command with data of 8 to disable the knob. If you then send another Set Device Mode command with data of 64 to enable message IDs, you will re-enable the knob since bit_3 in the 2nd instruction is 0.

Most devices have a default mode setting of 0 (all bits are 0).

Bit_n 2^n Description


bit_0 1 Disable Auto-reply
A value of 1 disables ALL replies except those to “return” commands. The default value is 0.

See Set Auto-Reply Disabled Mode (Cmd 101)

bit_1 2 Reserved
bit_2 4 Reserved
bit_3 8 Disable Knob
A value of 1 disables knob functionality. All knob turns and depresses are ignored. The default value is 0.

See Set Knob Disabled Mode (Cmd 107)

bit_4 16 Enable Move Tracking
A value of 1 enables the Move Tracking response during move commands. The device will return its position periodically when a move command is executed. The Disable Auto-Reply option above takes precedence over this option. The default value is 0.

See Set Move Tracking Mode (Cmd 115)

bit_5 32 Disable Manual Move Tracking
A value of 1 disables the Manual Move Tracking response during manual moves. The Disable Auto-Reply option above takes precedence over this option. The default value is 0.

See Set Manual Move Tracking Disabled Mode (Cmd 116)

bit_6 64 Enable Message Ids
A value of 1 enables Message Ids. When message id is enabled, only bytes 3 through 5 in the command requests and responses are used for data. Byte 6 is used as an id byte that the user can set to any value they wish. When the device receives a command request with a certain id byte, it will reply with a command response containing the same id byte. This allows the user application to monitor communication packets individually to implement error detection and recovery. The default value is 0.

See Set Message Id Mode (Cmd 102)

bit_7 128 Home Status
This value indicates whether a device has been homed and has a valid position reference. This bit is set to 0 automatically on power-up or reset. It is set automatically when the device is homed or when the position is set using command #45.

See Set Home Status (Cmd 103)

bit_8 256 Disable Auto-Home
A value of 1 prevents the home sensor from triggering a homing procedure during a movement command. This allows rotational devices to move multiple revolutions without re-triggering the homing mechanism.

See Set Auto-Home Disabled Mode (Cmd 105)

bit_9 512 Reverse Knob
A value of 1 reverses the direction of the travel when the knob is turned. The default value is 0.

See Set Knob Direction (Cmd 108)

bit_10 1,024 Reserved
bit_11 2,048 Reserved
bit_12 4,096 Set Home Switch Logic
A value of 1 indicates that the peripheral is installed with an active high home sensor. Most Zaber peripherals have active low home sensors. For peripherals installed with active high home sensors, this bit must be changed from the default setting in order for the home sensor to function correctly. Damage to the home sensor or actuator may result if this bit is set improperly.

See Set Home Sensor Type (Cmd 104)

NOTE: This setting in only valid on A-Series controllers, setting this bit on other devices will result in an error.

bit_13 8,192 Reserved
bit_14 16,384 Reserved
bit_15 32,768 Reserved


Set Home Speed - Cmd 41

Instruction Name Set Home Speed
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 41
Command Type Setting
Command Data Speed
Reply Data Speed
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the speed at which the device moves when using the "Home" command.

When a home instruction is issued, the device will accelerate at a rate determined by the acceleration setting up to the speed determined by this command. The home speed can be set independently of target speed (command 42) so that for stages that move very fast, issuing the home command would use a slower home speed and prevent the stage from crashing into the home position.

For a spreadsheet that can be used to calculate speed setting values for any product see http://www.zaber.com/documents/ZaberSpeedSetting.xls. Alternatively you may use the formulas below.

Actual Speed

= Data / 1.6384 * M mm/s or deg/s
= Data / 1.6384 microsteps/s
= Data / 1.6384 / R steps/s
= Data / 1.6384 / (R x S) * 60 revolutions/min Motor rpm
= Data / 1.6384 * L / (R x S) mm/s Linear devices only

where:

  • Data is the value of the command data
  • R (microsteps/step) is the microstep resolution (command 37)
  • S (steps/revolution) is the number of steps per motor revolution
  • M (mm or deg) is the microstep size
  • L (mm or deg) is the distance of linear motion per motor revolution

Valid data values are from 1 to (16384*R). Note that the maximum speed possible is independent of the resolution.

Refer to product specifications for the distance corresponding to a single microstep or revolution.

For example, if a motor has 200 steps per revolution (S = 200), used with default resolution (R = 64), and Data is 153600, then the motor will move at a speed of 720 revolutions per minute.

Set Target Speed - Cmd 42

Instruction Name Set Target Speed
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 42
Command Type Setting
Command Data Speed
Reply Data Speed
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the speed at which the device moves when using the "Move Absolute" or "Move Relative" commands.

When a move absolute or move relative instruction is issued, the device will accelerate at a rate determined by the acceleration setting up to the speed determined by this command.

The target velocity may be changed on-the-fly even when the device is in the middle of a move. The device will automatically adjust the velocity, but still target the final position specified in the original move.

For a spreadsheet that can be used to calculate speed setting values for any product see http://www.zaber.com/documents/ZaberSpeedSetting.xls. Alternatively you may use the formulas below.

Actual Speed

= Data / 1.6384 * M mm/s or deg/s
= Data / 1.6384 microsteps/s
= Data / 1.6384 / R steps/s
= Data / 1.6384 / (R x S) * 60 revolutions/min Motor rpm
= Data / 1.6384 * L / (R x S) mm/s Linear devices only

where:

  • Data is the value of the command data
  • R (microsteps/step) is the microstep resolution (command 37)
  • S (steps/revolution) is the number of steps per motor revolution
  • M (mm or deg) is the microstep size
  • L (mm or deg) is the distance of linear motion per motor revolution

Valid data values are from 1 to (16384*R). Note that the maximum speed possible is independent of the resolution.

Refer to product specifications for the distance corresponding to a single microstep or revolution.

For example, if a motor has 200 steps per revolution (S = 200), used with default resolution (R = 64), and Data is 153600, then the motor will move at a speed of 720 revolutions per minute.

Set Acceleration - Cmd 43

Instruction Name Set Acceleration
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 43
Command Type Setting
Command Data Acceleration
Reply Data Acceleration
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the acceleration used by the movement commands.

When a movement command is issued, the device will accelerate and decelerate at a rate determined by this command. The acceleration may be changed on-the-fly even when the device is in the middle of a move.

It is possible to set acceleration and deceleration separately in A-Series devices. In which case, use Set Acceleration Only (Cmd 113) and Set Deceleration Only (Cmd 114) instead. When acceleration and deceleration values are different, Return Setting(Data = Set Acceleration) returns the acceleration value only.

To determine the acceleration that will result from a given data value, the following formulas may be used:

Actual Acceleration
= 10000 * Data * M / 1.6384 mm/s^2 or deg/s^2
= 10000 * Data / 1.6384 microsteps/s^2
= 10000 * Data / R / 1.6384 steps/s^2

Where:

  • Data is the value specified in the Command Data
  • M (mm or deg) is the microstep size
  • R is the microstep resolution set in command #37 (microsteps/step)

The maximum value is 32767.

Set Maximum Position - Cmd 44

Instruction Name Set Maximum Position
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 44
Command Type Setting
Command Data Range
Reply Data Range
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the maximum position the device is allowed to travel to.

Use this command to limit the range of travel to a value other than the default. Exercise caution when using this command, since it is possible to set the range to a value greater than the physical limits of the device.

A device within range of travel is not allowed to move above its Maximum Position. If the device Current Position is out of range and above Maximum Position, the device is not allowed to move in the positive direction.

The default Maximum Position is the full travel length of the device. Valid values can be any number from -1000000000 to 1000000000.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Current Position - Cmd 45

Instruction Name Set Current Position
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 45
Command Type Setting
Command Data New Position
Reply Data New Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence Volatile
Summary Sets the device internal position counter.

This command override the internal position counter with a new position value specified by the user. The position data is volatile and will not persist after power-down or reset.

Stepper motor phase remains unchanged when position is updated.

Do not use this command to preserve the position counter over power cycles. To restore the position counter without homing the device, Park the stage before power down and Unpark the stage when power is restored. See Set Park State (Cmd 65) for details.

Set Home Offset - Cmd 47

Instruction Name Set Home Offset
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 47
Command Type Setting
Command Data Offset
Reply Data Offset
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the the new "Home" position which can then be used when the Home command is issued.


When the home command is issued, the device will retract until the home sensor is triggered, then move forward until the home sensor is no longer triggered, then move forward by the Home Offset value (in microsteps) and resets the internal position (to 0 for most devices).

Note that the home offset command also changes the range settings of the device. For example, if the initial Home Offset is 0 and the Minimum and Maximum Position are 0 and 500,000 respectively, and the user changes the Home Offset to 70,000, then the Minimum and Maximum Position are automatically adjusted to -70,000 and 430,000 respectively. However, changing the Minimum/Maximum Position does not affect the home offset.

When a new Home Offset is specified, Minimum and Maximum Positions are adjusted to provide the same range. The device is able to move in the same travel range as before.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Alias Number - Cmd 48

Instruction Name Set Alias Number
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 48
Command Type Setting
Command Data Alias Number
Reply Data Alias Number
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets an alternate device number for a device.

This setting specifies an alternate device number for a device (in addition to its actual device number). By setting several devices to the same alias number, you can control a group of devices with a single instruction. When you send an instruction to an alias number, all devices with that alias number will execute the instruction and reply using their actual device numbers. To remove an alias, simply set the device's alias number to zero. Valid alias numbers are 0 to 99 for version 6.05, and 0 to 254 otherwise. To avoid confusion, it is best to choose an alias greater than the number of devices connected.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Return Device Id - Cmd 50

Instruction Name Return Device Id
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 50
Command Type Read-Only Setting
Command Data Ignored
Reply Data Device Id
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the id number for the type of device connected.

See the Zaber support web site for a table of device ids for all Zaber products.

Return Firmware Version - Cmd 51

Instruction Name Return Firmware Version
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 51
Command Type Read-Only Setting
Command Data Ignored
Reply Data Version
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the firmware version installed on the device.

A decimal is assumed before the last two digits. For example, 502 indicates firmware version 5.02.

Return Power Supply Voltage - Cmd 52

Instruction Name Return Power Supply Voltage
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 52
Command Type Read-Only Setting
Command Data Ignored
Reply Data Voltage
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the voltage level of the device's power supply.

A decimal is assumed before the last digit. For example, a value of 127 indicates 12.7 V. Note that the internal voltage measurement is not very accurate. Don't be alarmed if the indicated voltage is slightly different from your measurements.

Return Setting - Cmd 53

Instruction Name Return Setting
Applies to All Zaber devices
Firmware Version 5.00 and up
Command Number 53
Command Type Command
Command Data Setting Number
Reply Data Setting Value
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the current value of the setting specified in the Command Data.

Valid command data values are the command numbers of any "Set..." instruction. The device will reply using the command number of the specified setting (as if a command to change the setting had just been issued) but the setting will not be changed.

For example, command #48 is the "Set Alias" instruction. Therefore if you wish to return the current value of the alias number, simply send the Return Setting instruction with data of 48. The device will reply with command #48 and data equal to the setting value.

Since firmware version 5.21, this command also accepts the command numbers of any "Return..." instruction, such as command #50 "Return Device Id".

Return Status - Cmd 54

Instruction Name Return Status
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 54
Command Type Read-Only Setting
Command Data Ignored
Reply Data Status
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Returns the current status of the device.

Possible status codes are as follows:

  • 0 - idle, not currently executing any instructions
  • 1 - executing a home instruction
  • 10 - executing a manual move in Velocity Mode (i.e. the manual control knob is turned)
  • 11 - executing a manual move in Displacement Mode (i.e. the manual control knob is turned)
  • 20 - executing a move absolute instruction
  • 21 - executing a move relative instruction
  • 22 - executing a move at constant speed instruction
  • 23 - executing a stop instruction (i.e. decelerating)
  • 65 - device is parked (FW 6.02 and up only. FW 6.01 returns 0 when parked)

Echo Data - Cmd 55

Instruction Name Echo Data
Applies to All Zaber devices
Firmware Version 5.04 and up
Command Number 55
Command Type Command
Command Data Data
Reply Data Data
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Echoes back the same Command Data that was sent.

This command is useful for testing communication, similar to a network "ping".

Return Current Position - Cmd 60

Instruction Name Return Current Position
Applies to All motorized devices
Firmware Version 5.00 and up
Command Number 60
Command Type Read-Only Setting
Command Data Ignored
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Returns the current absolute position of the device in microsteps.

This is equivalent to issuing a Return Setting (#53) command with a command data value of 45 (Set Current Position).

Set Park State - Cmd 65

Instruction Name Set Park State
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 65
Command Type Setting
Command Data Park State
Reply Data Park State
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Park or unpark the device.

This command allows a device to maintain its position reference over a power cycle without requiring homing. This feature is useful when a device needs to be powered down but a Home procedure is not desired at next power up.

Parking
To Park the device, issue command 65 with data value of 1. Device will store its internal position settings and shut off all current to the motor. Device can now be powered down.
Unparking
To Unpark the device, issue command 65 with data value of 0. Device will resume its previous internal settings and enable current to the motor. Current Position is now set to that when device was parked.
  • When a device is parked, no current is applied to the motor. Any movement command except Home results in an error response.
  • A parked device can be identified by its power LED (green) fading in and out slowly.
  • A parked device returns 65 upon Return Status (Cmd 54).
    • Exception: Firmware version 6.01 returns 0.
  • A parked device becomes unparked when one of the following commands is received:
1. Unpark (Set Park State (Cmd 65))
2. Home (Cmd 1)
3. Restore Settings (Cmd 36)
  • If the device is displaced while parked, the device should be parked again before powering down.
  • If the device is displaced since last parked or during powered off, a Homing procedure is required at the next start up. Unparking a displaced device will not resume correct settings.
  • If a device is not parked, an unpark command has no effect on the device.

Set Peripheral Id - Cmd 66

Instruction Name Set Peripheral Id
Applies to A-Series motor controllers
Firmware Version 6.xx
Command Number 66
Command Type Setting
Command Data Peripheral Id
Reply Data Peripheral Id
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Set default values for a specific peripheral device.

This command assigns the given Peripheral ID to the controller and loads the default settings for the assigned Peripheral Id. Once a Peripheral Id is assigned, Restore Settings (Cmd 36) with command data 0 can be issued to restore to default settings of the assigned Peripheral Id without needing to specify the Id.

If 0 is given instead of a Peripheral Id, the controller will restore to its safe-mode settings. At safe mode, the peripheral will be set to 10% maximum running current and no hold current. Any previously stored Peripheral Id will be erased. This is a relatively safe setting that prevents damage to any motor connected to the controller, but as a result, you may not be able to run a motor with the controller set to this condition. You can either issue the Set Peripheral Id instruction with one of the Peripheral Ids, or you can adjust each relevant setting independently.

The Set Peripheral Id instruction should initially be issued with a Peripheral Id corresponding to the specific motor or motorized device connected. This will set the motor controller to the factory default settings for the motorized peripheral in question. For example, if a LSA10A-T4 is connected to a controller, the user can issue the Set Peripheral Id instruction with a Peripheral Id of 48402. This will automatically set all non-volatile parameters (range, mode, running and hold currents, etc) to values that will work with the LSA10A-T4. For a list of Peripheral Ids and their corresponding settings, see http://www.zaber.com/support/?tab=Peripheral%20ids.

Note: Sending Set Peripheral Id with a peripheral Id value is equivalent to sending Restore Settings (Cmd 36) with a peripheral Id value. However, Set Peripheral Id with data 0 puts controller into safe-mode settings, whereas Restore Settings with data 0 causes the controller to load default values for the assigned peripheral Id.

Set Auto-Reply Disabled Mode - Cmd 101

Instruction Name Set Auto-Reply Disabled Mode
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 101
Command Type Setting
Command Data Auto-Reply Disabled Mode
Reply Data Auto-Reply Disabled Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Disables all command auto-replies from the device.

Use this command to disable auto-reply. When the auto-reply is disabled, device will only reply to “Return” commands.

0 - Auto-Reply Enabled (Default)
Device replies to all commands
1 - Auto-Reply Disabled
Device only replies to "Return" commands

This setting is identical to bit 0 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Message Id Mode - Cmd 102

Instruction Name Set Message Id Mode
Applies to A-Series devices
Firmware Version 6.xx
Command Number 102
Command Type Setting
Command Data Message Id Mode
Reply Data Message Id Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Enable message ids in device communication.

When message id is enabled, only bytes 3 through 5 in the command requests and responses are used for data. Byte 6 is used as an id byte that the user can set to any value they wish. When the device receives a command request with a certain id byte, it will reply with a command response containing the same id byte. This allows the user application to monitor communication packets individually to implement error detection and recovery.

0 - Message Id Disabled (Default)
All 4 data bytes are used for data
1 - Message Id Enabled
Only bytes 3 through 5 are used for data. Byte 6 is used as an id byte. Device always responds with a message id corresponding to that in the command request.
  • A device will not reply with the same message id more than once, unless user issues two commands with identical message id.
  • For any reply that is not a result of a user request, message id 0 is used. (e.g. Manual Move Tracking)
  • NOTE: When message id is enabled, data range is limited to 3 bytes (24 bits). Thus, commands with data value more than 24 bit range might not behave properly.

This setting is identical to bit 6 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Example
requests -> (id 01) Move Absolute 10000 [device starts to move]
requests -> (id 02) Return Status [request device status]
<- responds (id 02) Return Status ( 20 ) [device replies with message id corresponding to the request (id 02)]
<- responds (id 01) Move Absolute 10000 [when movement is completed, device replies with id 01 which corresponds to the request.]

Set Home Status - Cmd 103

Instruction Name Set Home Status
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 103
Command Type Setting
Command Data Home Status
Reply Data Home Status
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Sets whether a device has been homed.

This value indicates whether a device has been homed and has a valid position reference.

  • This value is set to 0 automatically on power-up or reset
  • This value is set to 1 automatically when the device is homed or when the position is set using command #45

This setting is identical to bit 7 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

Set Home Sensor Type - Cmd 104

Instruction Name Set Home Sensor Type
Applies to A-Series motor controllers
Firmware Version 6.xx
Command Number 104
Command Type Setting
Command Data Home Sensor Type
Reply Data Home Sensor Type
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Configures the home sensor type of the device.

Use this command to select the home sensor output level the firmware should interpret as a home sensor trigger. Most Zaber peripherals have active low home sensors. For peripherals installed with active high home sensors, this setting must be changed from the default value in order for the home sensor to function correctly. Damage to the home sensor or actuator may result if this value is set improperly.

0 - Active low home sensor (Default)
Home sensor outputs low voltage level when triggered
1 - Active high home sensor
Home sensor outputs high voltage level when triggered

This setting is identical to bit 12 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Auto-Home Disabled Mode - Cmd 105

Instruction Name Set Auto-Home Disabled Mode
Applies to A-Series motor controllers
Firmware Version 6.xx
Command Number 105
Command Type Setting
Command Data Auto-Home Disabled Mode
Reply Data Auto-Home Disabled Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Prevents home sensor from triggering a homing procedure during movement commands.

Use this command to disable auto-home. During a movement command with auto-home enabled, a device homes itself whenever the home sensor is triggered. The home sensor is ignored if auto-home is disabled. This allows rotational devices to move multiple revolutions without re-triggering the homing mechanism.

0 - Auto-Home Enabled (Default)
Device runs homing procedure whenever home sensor is triggered
1 - Auto-Home Disabled
Device responses to home sensor only during a Home command

This setting is identical to bit 8 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Minimum Position - Cmd 106

Instruction Name Set Minimum Position
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 106
Command Type Setting
Command Data Minimum Position
Reply Data Minimum Position
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the minimum position the device is allowed to reach.

A device within range of travel is not allowed to move below its Minimum Position. If the device Current Position is out of range and below Minimum Position, the device is not allowed to move in the negative direction. The default Minimum Position in most devices is 0.

Use this command to limit the range of travel to a value other than the default. Exercise caution when using this command, since it is possible to set the range to a value greater than the physical limits of the device.

Valid values can be any number from -1000000000 to 1000000000.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Disabled Mode - Cmd 107

Instruction Name Set Knob Disabled Mode
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 107
Command Type Setting
Command Data Knob Disabled Mode
Reply Data Knob Disabled Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Enables or disables manual movement control.

Use this command to disable knob functionality. When the knob is disabled, all knob turns and depresses are ignored.

0 - Knob Enabled (Default)
Knob functions normally.
1 - Knob Disabled
Knob interaction is ignored.

This setting is identical to bit 3 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Direction - Cmd 108

Instruction Name Set Knob Direction
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 108
Command Type Setting
Command Data Knob Direction
Reply Data Knob Direction
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes direction of movement when knob is turned.

Use this command to reverse the device movement direction when the knob is turned.

0 - Normal (Default)
Clockwise knob turn results in movement in the positive direction.
1 - Reversed
Clockwise knob turn results in movement in the negative direction.

This setting is identical to bit 9 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Movement Mode - Cmd 109

Instruction Name Set Knob Movement Mode
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 109
Command Type Setting
Command Data Movement Mode
Reply Data Movement Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the movement mode used when the manual knob is turned.

Use this command to set the desired manual movement mode.

0 - Velocity Mode (Default)
Each detent of a knob turn moves the device at an increasing speed. There are 16 speeds in each direction. The velocity profile and the maximum speed can be set via Set Knob Velocity Profile (Cmd 112) and Set Knob Velocity Scale (Cmd 111).
1 - Displacement Mode
Each detent of a knob turn moves the device for a fixed number of microsteps. The device stops if the number of microsteps left in the maximum range is less than the requested move command number. The number of microsteps per knob detent can be set via Set Knob Jog Size (Cmd 110).

NOTE: User can also quickly toggle between the two modes by depressing and holding the knob for 1 second.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Jog Size - Cmd 110

Instruction Name Set Knob Jog Size
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 110
Command Type Setting
Command Data Jog Size
Reply Data Jog Size
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the distance to travel for knob manual movement in displacement mode.

Use this command to set the number of microsteps per knob detent when the knob is turned in displacement mode.

Each detent of a knob turn moves the device for a fixed number of microsteps. The device stops if the number of microsteps left in the max/min range is less than the requested move command number.

The default setting for most devices sets a value such that one full cycle of the knob will result in one revolution of the stepper motor.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Velocity Scale - Cmd 111

Instruction Name Set Knob Velocity Scale
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 111
Command Type Setting
Command Data Maximum Speed
Reply Data Maximum Speed
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the maximum speed for knob manual movement in velocity mode.

Use this command to set the desired maximum speed for a knob manual move in velocity mode. This speed corresponds to the 16th detent on the knob. Intermediate speeds are determined by the velocity profile (See Set Knob Velocity Profile (Cmd 112)).

This value only applies when the device has been homed. Before homing, the maximum speed when using a knob manual move in velocity mode is the Home Speed (See Set Home Speed (Cmd 41)).

The default velocity scale is identical to that of Set Target Speed (Cmd 42).

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Knob Velocity Profile - Cmd 112

Instruction Name Set Knob Velocity Profile
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 112
Command Type Setting
Command Data Profile Number
Reply Data Profile Number
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the relationship between knob turn and the velocity of the device for knob manual movement in velocity mode.

Use this command to set how device speed should vary when the knob is turned in velocity mode. There are 16 knob detents for 16 speeds in each direction. The 16th detents are scaled relative to the maximum speed set by Set Knob Velocity Scale (Cmd 111).

Three velocity profiles are available:

  • 1 – Linear
  • 2 – Quadratic (Default)
  • 3 – Cubic

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Acceleration Only - Cmd 113

Instruction Name Set Acceleration Only
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 113
Command Type Setting
Command Data Acceleration
Reply Data Acceleration
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets only the acceleration used by the movement commands, leaving deceleration value unchanged.

When a movement command is issued, the device will accelerate at a rate determined by this command up to a maximum speed determined by the target velocity.

This command only changes the acceleration value but not the deceleration value. Use Set Deceleration Only (Cmd 114) to change only the deceleration value, or use Set Acceleration (Cmd 43) to set both acceleration and deceleration at the same time.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Deceleration Only - Cmd 114

Instruction Name Set Deceleration Only
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 114
Command Type Setting
Command Data Deceleration
Reply Data Deceleration
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets only the deceleration used by the movement commands, leaving acceleration value unchanged.

When a movement command is issued, the device will decelerate to a stop at a rate determined by this command.

This command only changes the deceleration value but not the acceleration value. Use Set Acceleration Only (Cmd 113) to change only the acceleration value, or use Set Acceleration (Cmd 43) to set both acceleration and deceleration at the same time.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Move Tracking Mode - Cmd 115

Instruction Name Set Move Tracking Mode
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 115
Command Type Setting
Command Data Move Tracking Mode
Reply Data Move Tracking Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Enable Move Tracking messages.

Enable Move Tracking to receive position responses during move commands. The device will return its position periodically when a move command is executed.

0 - Move Tracking Disabled (Default)
Device does not send Move Tracking messages while executing a move command.
1 - Move Tracking Enabled
Device sends Move Tracking responses periodically while executing a move command.

This setting is identical to bit 4 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Example

Move Tracking Disabled

requests -> Move Absolute 100000
<- responds Move Absolute 100000 [ Device arrives at target position ]

Move Tracking Enabled

requests -> Move Absolute 100000
<- responds Move Tracking (Position 19892)
<- responds Move Tracking (Position 43320)
<- responds Move Tracking (Position 66767)
<- responds Move Tracking (Position 90195)
<- responds Move Absolute 100000 [ Device arrives at target position ]

Set Manual Move Tracking Disabled Mode - Cmd 116

Instruction Name Set Manual Move Tracking Disabled Mode
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 116
Command Type Setting
Command Data Manual Move Tracking Disabled Mode
Reply Data Manual Move Tracking Disabled Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Disable Manual Move Tracking messages.

Disable Move Tracking to inhibit position responses during manual movements when using the knob in velocity mode. When manual move tracking is enabled, the device will return its position periodically when a manual move command is executed.

0 - Manual Move Tracking Enabled (Default)
Device sends Manual Move Tracking messages periodically when knob is used in Velocity mode.
1 - Manual Move Tracking Disabled
Device does not send Manual Move Tracking response.

This setting is identical to bit 5 of Set Device Mode (Cmd 40). Changes to either this setting or the device mode will be reflected in the other.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Example

Manual Move Tracking Enabled

*** Knob in Velocity Mode ***
*** Knob turned *** [ Device begins to move ]
<- responds Manual Move Tracking (Position 19892)
<- responds Manual Move Tracking (Position 43320)
<- responds Manual Move Tracking (Position 66767)
*** Knob turned to Stop position ***
<- responds Manual Move Tracking (Position 90195) [ Device stopped ]

Manual Move Tracking Disabled

*** Knob in Velocity Mode ***
*** Knob turned *** [ Device begins to move ]
*** Knob turned to Stop position ***
<- responds Manual Move Tracking (Position 90195) [ Device stopped ]

Set Move Tracking Period - Cmd 117

Instruction Name Set Move Tracking Period
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 117
Command Type Setting
Command Data Tracking Period in ms
Reply Data Tracking Period in ms
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the update interval between Move Tracking and Manual Move Tracking messages.

Use this command to configure the interval between each Move Tracking or Manual Move Tracking responses. The value is specified in milliseconds and the default setting is 250ms.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Closed-Loop Mode - Cmd 118

Instruction Name Set Closed-Loop Mode
Applies to A-Series motorized closed-loop devices
Firmware Version 6.xx
Command Number 118
Command Type Setting
Command Data Closed-Loop Mode
Reply Data Closed-Loop Mode
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the closed-loop mode for encoder embedded devices.

This command configures the behaviour of a closed-loop device.

0 - Disabled
Encoder input is ignored. Device behaves identically to other open-loop devices.
1 - Passive
Device sends Slip responses whenever slipping is detected.
No adjustment or correction is made.
2 - Stall Detection
Device sends Slip responses whenever slipping is detected.
If stalling is detected, device will stop after a stall timeout. See Set Stall Timeout (Cmd 120).
No adjustment or correction is made.
3 - Position Correction (Default)
Device sends Slip responses whenever slipping is detected.
Current Position is updated to actual location whenever a slip is detected.
If stalling is detected, device will stop after a stall timeout. See Set Stall Timeout (Cmd 120).
4 - Stall Recovery
Device sends Slip responses whenever slipping is detected.
Current Position is updated to actual location whenever a slip is detected.
If stalling is detected, device will stop and resume after a stall timeout. See Set Stall Timeout (Cmd 120).
5 - Stall and Displace Recovery
Device sends Slip responses whenever slipping is detected.
Current Position is updated to actual location whenever a slip is detected.
If stalling is detected, device will stop and resume after a stall timeout. See Set Stall Timeout (Cmd 120).
When a stationary device is forced out of position, device will restore to original location after a timeout. See Set Stall Timeout (Cmd 120).
6 - Low Tolerance Stall and Displacement Recovery (Experimental mode - potentially unstable)
This mode is identical to Mode 5, but device sensitivity is tuned up to lower stall occurrence and to boost operating thrust.
WARNING: This is an experimental mode and is potentially unstable. Do not use this mode when mounting sensitive equipment to your Zaber product. Use at your own risk. Zaber is not responsible for any damage resulting from this closed-loop mode.


Mode 0 1 2 3 (Default) 4 5
Name Disabled Passive Stall Detection Position Correction Stall Recovery Stall and Displace Recovery
Send responses while slipping*   Cmd 12* Cmd 12* Cmd 12* Cmd 12* Cmd 12*
Always update Current Position to actual location       Y Y Y
Stall: When device is stalling...     Stop** Stop** Stop & Resume** Stop & Resume**
Send response after device has stopped     Cmd 13 Cmd 13    
Displaced: When a stationary device is forced out of position...           Restore**
Send responses while being displaced*   Cmd 13* Cmd 13* Cmd 13* Cmd 13* Cmd 12*
Send response after location is recovered           Cmd 13
*Response period can be adjusted or turned off via Set Slip Tracking Period (Cmd 119)
**Duration before action is taken can be configured via Set Stall Timeout (Cmd 120)

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Slip Tracking Period - Cmd 119

Instruction Name Set Slip Tracking Period
Applies to A-Series motorized closed-loop devices
Firmware Version 6.xx
Command Number 119
Command Type Setting
Command Data Tracking Period in ms
Reply Data Tracking Period in ms
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the update interval between Slip Tracking messages.

Use this command to configure the interval between each Slip Tracking responses.

0 - Slip Tracking Disabled
Set to 0 to turn off slip tracking
1 to 65535 - Slip Tracking Enabled
Interval between Slip Tracking messages, in units of ms.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Stall Timeout - Cmd 120

Instruction Name Set Stall Timeout
Applies to A-Series motorized closed-loop devices
Firmware Version 6.xx
Command Number 120
Command Type Setting
Command Data Stall Timeout in ms
Reply Data Stall Timeout in ms
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Sets the timeout for stall and forced displacement conditions in closed-loop modes.

Use this command to configure the duration before an action is taken when a stall or forced displacement occurs. The value is specified in ms and the default is 500 ms.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Device Direction - Cmd 121

Instruction Name Set Device Direction
Applies to A-Series motor controllers
Firmware Version 6.xx
Command Number 121
Command Type Setting
Command Data Device Direction
Reply Data Device Direction
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes direction of device movement.

Use this command to reverse the device movement direction if the device is homing in the wrong direction. When set to reversed mode, motor output and encoder input will increment in the opposite direction. This is equivalent to swapping the phases on motor and encoder cables, or attaching a motor at the opposite end of the stage. Improperly setting this value may result in damage to the home sensor or actuator.

0 - Normal (Default)
Device operates in normal direction.
1 - Reversed
Device operates in reversed direction.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Baudrate - Cmd 122

Instruction Name Set Baudrate
Applies to A-Series devices
Firmware Version 6.06 and up
Command Number 122
Command Type Setting
Command Data Baudrate
Reply Data Baudrate
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes the baudrate of the RS232 communications

This command sets the baudrate in use for communications. The device will reply on the current baudrate and then change to the new setting after the communication lines have been idle for 500ms.

Valid baudrates are:

  • 9600.
  • 19200.
  • 38400.
  • 57600.
  • 115200.

All other serial parameters are 8 bits, 1 stop bit, No parity and no flow control.

To set protocol and RS232 baudrate at the same time, use the Convert To Ascii (Cmd 124) command.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Set Protocol - Cmd 123

Instruction Name Set Protocol
Applies to A-Series devices
Firmware Version 6.06 and up
Command Number 123
Command Type Setting
Command Data Protocol
Reply Data Protocol
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes the communications protocol in use.

This command sets the protocol in use for communications. The device will reply in the current protocol and then change to the new setting after the communication lines have been idle for 500ms.

Valid protocols are:

To set protocol and RS232 baudrate at the same time, use the Convert To Ascii (Cmd 124) command.

This setting is stored in non-volatile memory and will persist after power-down or reset.

Convert To Ascii - Cmd 124

Instruction Name Convert To Ascii
Applies to A-Series devices
Firmware Version 6.06 and up
Command Number 124
Command Type Command
Command Data Baudrate
Reply Data Baudrate
Safe to retry? Yes
Returns Current Position? No
Persistence Non-Volatile
Summary Changes the RS232 baudrate and sets the communications protocol to ASCII

This command sets the baudrate in use for communications and sets the protocol to ASCII Protocol. The device will reply on the current baudrate and protocol, then change to the new settings after the communication lines have been idle for 500ms.

Valid baudrates are:

  • 9600.
  • 19200.
  • 38400.
  • 57600.
  • 115200.

All other serial parameters are 8 bits, 1 stop bit, No parity and no flow control.

The settings are stored in non-volatile memory and will persist after power-down or reset.

Error - Cmd 255

Instruction Name Error
Applies to A-Series devices
Firmware Version 6.xx
Command Number 255
Command Type Reply
Command Data n/a
Reply Data Error Code
Safe to retry? Yes
Returns Current Position? No
Persistence n/a
Summary Indicates to the user that an error has occurred.

This reply indicates that an error has occurred. The error code returned in the data indicates the type of error. The device may send an error code as a reply to an invalid instruction, or it may autonomously send an error code as a triggered reply (i.e. not in response to an instruction). The error code is typically the command number of the instruction that caused the error, but not always.

Error Codes
Code Name Description
1 Cannot Home Home - Device has traveled a long distance without triggering the home sensor. Device may be stalling or slipping.
2 Device Number Invalid Renumbering data out of range.
5 Address Invalid Read Register - Register address invalid.
14 Voltage Low Power supply voltage too low.
15 Voltage High Power supply voltage too high.
18 Stored Position Invalid The position stored in the requested register is no longer valid. This is probably because the maximum range was reduced.
20 Absolute Position Invalid Move Absolute - Target position out of range.
21 Relative Position Invalid Move Relative - Target position out of range.
22 Velocity Invalid Constant velocity move. Velocity out of range.
36 Peripheral Id Invalid Peripheral Id is invalid or not supported.
37 Resolution Invalid Microstep resolution not supported.
38 Run Current Invalid Run current out of range.
39 Hold Current Invalid Hold current out of range.
40 Mode Invalid Set Device Mode - one or more of the mode bits is invalid.
41 Home Speed Invalid Home speed out of range. The range of home speed is determined by the resolution.
42 Speed Invalid Target speed out of range. The range of target speed is determined by the resolution.
43 Acceleration Invalid Target acceleration out of range.
44 Maximum Position Invalid Maximum position out of range.
45 Current Position Invalid Current position out of range.
47 Offset Invalid Home offset out of range.
48 Alias Invalid Alias out of range.
50 Device Id Unknown The device id is not included in the firmware's list.
53 Setting Invalid Return Setting - data entered is not a valid setting command number. Valid data are the command numbers of any "Set..." or "Return..." instructions.
64 Command Invalid Command number not valid in this firmware version.
65 Park State Invalid Set Park State - State must be 0 or 1, or device cannot park because it is in motion.
67 Temperature High High temperature is detected inside device. Device may be overheating.
101 Auto Reply Disabled Mode Invalid Mode must be 0 or 1.
102 Message Id Mode Invalid Mode must be 0 or 1.
103 Home Status Invalid Status must be 0 or 1.
104 Home Sensor Type Invalid Type must be 0 or 1.
105 Auto-Home Disabled Mode Invalid Mode must be 0 or 1.
106 Minimum Position Invalid Minimum position out of range.
107 Knob Disabled Mode Invalid Mode must be 0 or 1.
108 Knob Direction Invalid Direction must be 0 or 1.
109 Knob Movement Mode Invalid Mode must be 0 or 1.
111 Knob Velocity Scale Invalid Maximum knob speed out of range. The range of valid speed is determined by the resolution.
112 Knob Velocity Profile Invalid Profile must be 1 (Linear), 2 (Quadratic), or 3 (Cubic).
113 Acceleration Only Invalid Acceleration out of range.
114 Deceleration Only Invalid Deceleration out of range.
115 Move Tracking Mode Invalid Mode must be 0 or 1.
116 Manual Move Tracking Disabled Mode Invalid Mode must be 0 or 1.
117 Move Tracking Period Invalid Valid range is 10 - 65535.
118 Closed-Loop Mode Invalid Valid modes are 0-6.
119 Slip Tracking Period Invalid Valid range is 0(Off), 10 - 65535.
120 Stall Timeout Invalid Valid range is 0 - 65535.
121 Device Direction Invalid Direction must be 0 or 1.
122 Baudrate Invalid Set Baudrate - Value not supported.
123 Protocol Invalid Set Protocol - Value not supported.
124 Baudrate or Protocol Invalid Set Baudrate and ASCII Protocol - Value not supported.
255 Busy Another command is executing and cannot be pre-empted. Either stop the previous command or wait until it finishes before trying again.
701 Register Address Invalid Write Register - Register address invalid.
702 Register Value Invalid Write Register - Value out of range.
1600 Save Position Invalid Save Current Position register out of range (must be 0-15).
1601 Save Position Not Homed Save Current Position is not allowed unless the device has been homed.
1700 Return Position Invalid Return Stored Position register out of range (must be 0-15).
1800 Move Position Invalid Move to Stored Position register out of range (must be 0-15).
1801 Move Position Not Homed Move to Stored Position is not allowed unless the device has been homed.
4001 Bit 1 Invalid Set Device Mode - bit 1 is reserved in this device and must be 0.
4002 Bit 2 Invalid Set Device Mode - bit 2 is reserved in this device and must be 0.
4008 Disable Auto Home Invalid Set Device Mode - this is a linear actuator; Disable Auto Home is used for rotary actuators only.
4010 Bit 10 Invalid Set Device Mode - bit 10 is reserved in this device and must be 0.
4011 Bit 11 Invalid Set Device Mode - bit 11 is reserved in this device and must be 0.
4012 Home Switch Invalid Set Device Mode - this device has integrated home sensor with preset polarity; mode bit 12 cannot be changed by the user.
4013 Bit 13 Invalid Set Device Mode - bit 13 is reserved in this device and must be 0.
4014 Bit 14 Invalid Set Device Mode - bit 14 is reserved in this device and must be 0.
4015 Bit 15 Invalid Set Device Mode - bit 15 is reserved in this device and must be 0.
6501 Device Parked Device is currently parked. Use Set Park State or Home to unpark device before requesting a move.

Reply-Only Reference

In general, a Zaber device will reply to an instruction using the same command number as the instruction itself. However, there are occasions (such as when the user turns a manual control knob) when the device may transmit data without first receiving a request from the controlling computer. This type of reply may be considered to be a triggered reply as opposed to a requested reply. In this case the device uses a “reply-only” command number to distinguish the reply from those requested by the controlling computer. The meanings of these replies and their corresponding data are given below.

Move Tracking - Cmd 8

Instruction Name Move Tracking
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 8
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Indicates to the user that the device has been set to a position tracking mode and given a move instruction.

Move Tracking has been enabled (see Set Move Tracking Mode (Cmd 115)) and device has been given a move instruction. In this mode, the device sends this reply at a regular interval updating the current absolute position (in microsteps) during any move. The default interval is 250ms and can be configured with Set Move Tracking Period (Cmd 117).

Limit Active - Cmd 9

Instruction Name Limit Active
Applies to All motorized devices
Firmware Version 5.00 and up
Command Number 9
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary Indicates to the user that the device has reached one of the limits of travel.

This response from a device indicates that it reached one of the limits of travel (either the minimum position or maximum position). For example, if a device is sent a “move at constant speed” instruction, and it reaches the maximum range of travel then the device will reply with this command number and the maximum position reached as data.

Manual Move Tracking - Cmd 10

Instruction Name Manual Move Tracking
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 10
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary A reply that is sent when the manual control knob is turned.

Manual Move Tracking has been enabled (see Set Manual Move Tracking Disabled Mode (Cmd 116)) and the device has been moved manually (the knob is turned). In this mode, the device sends this reply at a regular interval updating the current absolute position (in microsteps) during any move. The default interval is 250ms and can be configured with Set Move Tracking Period (Cmd 117).

Manual Move - Cmd 11

Instruction Name Manual Move
Applies to A-Series motorized devices
Firmware Version 6.xx
Command Number 11
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary A knob manual move in Displacement mode has completed.

The device has been moved manually (the knob is turned).

Slip Tracking - Cmd 12

Instruction Name Slip Tracking
Applies to A-Series motorized closed-loop devices
Firmware Version 6.xx
Command Number 12
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary A reply that is sent when the device is slipping.

If slip tracking is enabled in the selected Closed Loop Mode (Cmd 118) and the device begins slipping, a reply will be sent. The device sends this reply at regular intervals until device is no longer slipping. Use Set Slip Tracking Period (Cmd 119) to configure the period that this reply is sent with or to disable the replies.

Unexpected Position - Cmd 13

Instruction Name Unexpected Position
Applies to A-Series motorized closed-loop devices
Firmware Version 6.xx
Command Number 13
Command Type Reply
Command Data n/a
Reply Data Position
Safe to retry? Yes
Returns Current Position? Yes
Persistence n/a
Summary A reply that is sent when the device stops at a position different from the requested location.

This message is received in one of the following scenarios:

  • The device has stalled and stopped (Closed-loop mode 2-3)
  • The stationary device is being forced out of position (Closed-loop mode 1-4)
  • The stationary device is forced out of position and has recovered to its original location (Closed-loop mode 5)

This command indicates that no further closed-loop action (stop/resume/restore) will be taken. It is sent only when the device is at rest (Return Status = 0). See Set Closed-Loop Mode (Cmd 118) for more details.


Troubleshooting A-Series Closed-Loop Motion Devices

Symptom Check
Nothing happens when I turn the knob either way. Manual control may have been disabled. Issue the Restore Settings (Cmd 36) command or enable the knob using Set Knob Disabled Mode (Cmd 107) with data 0.
Nothing happens when I send a move instruction. The device needs to be "homed" first. You must issue the Home (Cmd 1) command after power up to initialize the device.
Device does not move when sending an instruction or turning the knob. Check whether the green LED is fading in and out slowly. If so, the device is Parked. Send Set Park State (Cmd 65) with data 0 to unpark the device, or Home (Cmd 1) the device.
The device can move smoothly, but only moves for a short time then stops. Blue LED is flashing. If the blue LED is flashing and Slip Tracking or Unexpected Position responses are received, but the device is not actually slipping or stalling, the internal encoder counter may need to be re-initialized. Reset the device by power cycling it or sending Reset (Cmd 0) command, then re-initialize it with the Home (Cmd 1) command. Ground the device and avoid operating it under statically noisy environment.
The device is moving on its own, and is running against the ends of travel. The internal encoder counter may need to be re-initialized. Reset the device by power cycling it or sending Reset (Cmd 0) command, then re-initialize it with the Home (Cmd 1) command. Ground the device and avoid operating it under statically noisy environment.
Stall condition: The device makes noise but does not move. Blue LED is flashing. Try removing all external loads. If the device now extends and retracts normally, the problem is excessive load. Try to reduce the load and ensure the load is less than the maximum thrust. You can achieve higher thrust or torque by lowering the speed of the device using the Set Target Speed Instruction (command 42). If a device is stalled with no external load then it requires servicing.
There is no communication between the computer and my Zaber device, the yellow light does not come on or flash. There are several things you should check:
  1. Make sure you have the correct serial port selected. Try selecting other serial ports in the software. To ensure that your serial port is working properly, you can connect the send and receive pins of the serial adapter without connecting any devices. Use a wire to connect pins 2 and 6. Your software should echo a reply every time you send an instruction. You can view the pinout diagram here.
  2. Check the baud rate, hand shaking, parity, stop bit, etc. when configuring your serial communications software. The required settings are listed in the Control Through The RS-232 Serial Port section above. Also make sure that your serial port is not configured with a termination character (it often defaults to linefeed).
  3. Make sure there are no bent pins in the ends of all the data cables
  4. Make sure the power supply is working. The dc plug should measure 12 - 48 Vdc when it's not plugged into the device.
  5. Make sure the power plug is plugged in all the way. If your device has LEDs, the green LED should light.
  6. If your computer is a laptop running on batteries, try plugging in the power. Some laptops disable the serial ports when running on batteries.
  7. Make sure you do not have a null modem adapter or cable in the line.
  8. The serial to mini-din adapter comes in many varieties and many have different pin connections. Check the adaptor for continuity on the proper pins by consulting the adaptor pin-out diagram above.
  9. If you encounter the problem when trying to control the device with your own software, try using one of the demo programs from our website to verify that the hardware is functioning properly.
My device is behaving strangely. It responds to some commands as expected but not to others.
  1. Send a Restore Settings Instruction (command 36). A setting might have been inadvertently changed. If you have a T-MCA or T-CD series stepper motor controller, note that a data value should be entered with the Restore Default Settings Instruction corresponding to the Peripheral Id of the device you are controlling.
  2. Your computer might be set to Unicode. This is common for languages that use non-Latin based characters. Go to Control Panel/Regional and Language Options/Advanced. Select a language for non-unicode programs. This should be English or another Latin based character language.
The device is moving very slowly. It used to behave differently. The speed settings may have been changed inadvertently. Send a Restore Settings Instruction (command 36).
Green LED does not come on Check the A/C wall plug, the voltage adaptor and its connection to the device. If the power is coming over the data cable, check the mini din connector for bent or broken pins.
Green LED flashes at 2Hz The power supply voltage or temperature is out of range. It may either be too low or too high.
Green LED fades in and out slowly.

Device does not move when sending an instruction or turning the knob.

The device is Parked. Send Set Park State (Cmd 65) with data 0 to unpark the device, or Home (Cmd 1) the device.
The yellow light comes on briefly when sending a command, but the device does not move and does not return codes. Check baud rate, hand shaking, parity, stop bit, etc. Make sure that your software does not transmit any control characters like line feed, spaces or something else. The device numbers may not be what you think they are. Issue a renumber command, make sure that the computer does not transmit anything else while the devices renumber. Check that you transmit 6 bytes and that the device number and command are valid. If you encounter the problem when trying to control the device with your own software, try using one of the demo programs from our website to verify that the hardware
The red LED is on or flashing Device has lost its settings, or an error has occurred. Please contact technical support.
The device does not send replies but otherwise works. Auto-reply might have been disabled via Set Auto-Reply Disabled Mode (Cmd 101), try sending a Restore Settings (Cmd 36) command. If you encounter the problem when trying to control the device with your own software, try using a demo program from our website to verify that the hardware is functioning properly. Make sure that the receiving part of your code or commercial package is correct. Check baud rate, etc. Check connectors for bent or broken pins.
The device sometimes returns fewer than 6 bytes. This problem usually indicates a problem with the settings for your serial port. Some serial ports are set to automatically recognize and remove specific control characters such as carriage returns when they appear in the RS232 receive buffer. When this happens, it appears as though the device has not sent enough bytes, but really the controlling computer has just removed some before you could read them. You will need to change the serial port settings to fix the problem.
The device makes louder than normal noise during travel and is frequently slipping. This condition happens if the thrust needed is more than the thrust available from the device. Check that the force on the device is less than the maximum thrust. Check the voltage using the voltage command. Voltage less than the specified voltage for the device will reduce the device’s maximum thrust. Try a slower target velocity (command #42) as stepper motors produce more thrust when moving slowly. Lead screw conditions greatly affect the performance of linear motion devices. Dirt, damaged threads, no grease or too much grease will degrade performance and may contribute to a stall. A black residue appears on the lead screw after extended use. This can increase friction and reduce thrust. Clean the screw and re-grease it with a grease that does not degrade plastics. In general if you try to move a large payload or have a large static axial load (like lifting something vertically) you will have more problems. For vertical motion, the use of a counterweight, spring or rubber band can help reduce the static load and improve the performance of the device. The default value of the acceleration and target velocity are good for small to medium loads and medium speeds. For very light loads and higher speeds, or heavy loads at lower speeds, these parameters can be tuned. Trial and error is the best tuning technique.
The device has repeatability errors smaller than 4 full steps. If you're not skipping steps, friction or loose parts may still cause some variation when returning to a position.
The device does not behave as expected when software sends it a series of commands. It can be challenging to track down problems in a complicated script or other software. In addition to the standard techniques of debugging any software, it's also helpful to see exactly what bytes are being sent and received on the serial port. There are several tools available to display the raw data from a serial port, and stackoverflow.com has a list of some.

Warranty and Repair

For Zaber's policies on warranty and repair, please refer to the Ordering Polices

Standard products

Standard products are any part numbers that do not contain the suffix ENG followed by a 4 digit number. Most, but not all, standard products are listed for sale on our website. All standard Zaber products are backed by a one-month satisfaction guarantee. If you are not satisfied with your purchase, we will refund your payment minus any shipping charges. Goods must be in brand new saleable condition with no marks. Zaber products are guaranteed for one year. During this period Zaber will repair any products with faults due to manufacturing defects, free of charge.

Custom products

Custom products are any part numbers containing the suffix ENG followed by a 4 digit number. Each of these products has been designed for a custom application for a particular customer. Custom products are guaranteed for one year, unless explicitly stated otherwise. During this period Zaber will repair any products with faults due to manufacturing defects, free of charge.

How to return products

Customers with devices in need of return or repair should contact Zaber to obtain an RMA form which must be filled out and sent back to us to receive an RMA number. The RMA form contains instructions for packing and returning the device. The specified RMA number must be included on the shipment to ensure timely processing.

Email Updates

If you would like to receive our periodic email newsletter including product updates and promotions, please sign up online at www.zaber.com (news section). Newsletters typically include a promotional offer worth at least $100.

Contact Information

Contact Zaber Technologies Inc by any of the following methods:

Phone 1-604-569-3780 (direct)
1-888-276-8033 (toll free in North America)
Fax 1-604-648-8033
Mail #2 - 605 West Kent Ave. N., Vancouver, British Columbia, Canada, V6P 6T7
Web www.zaber.com
Email Please visit our website for up to date email contact information.

Appendix A: Default Settings

Please see the Zaber Support Page for default settings for this device.

Appendix B: Device Specifications

For complete device specifications for A-BLQ-E stages, please see our website.